RU EN

IKV-1-3-1 vibration measuring channel

IKV-1-3-1 vibration measuring channel

Designed to measure the vibration displacement range in diagnostic and emergency protection systems (EPS)*;

Consisting of a DS-x eddy-current transducer, an AS 131 transducer (in a connection box), or an AS 1312 transducer in a connector.

Can be used on compressor equipment;

Channel versions:

  • IKV-1-3-1 ver. A - DS-0 (DS-1, DS-2, DS-3) ver. 00 or 01 (with connector) eddy-current transducer, AS 131 converter, (connection box, TIK-BIS.121.0X0X safety barrier)**;
  • IKV-1-3-1 ver. B - DS-0 (DS-1, DS-2, DS-3) ver. 02 eddy-current transducer, AS 1312 converter in the socket, (connection box, TIK-BIS.121.0X0X safety barrier)**;
  • IKV-1-3-1 ver. E - DS-0 (DS-1, DS-2, DS-3) ver. 00 or 01 (with connector) eddy-current transducer, AS 131 converter (with display), (connection box, TIK-BIS.121.0X0X safety barrier)**.

Output signal type:

  • 4-20 mA (for ver. A, B, E);
  • 0-10 V (for ver. E);
  • digital output (for ver. E).

*By default, the channel is set to 40X metal

**Available as an option

Additional photos

Reviews

Related equipment

TIK-BIS
safety barriers

specifications

Metrological parameters

  • Conversion coefficient, mA/μm / Range of indications of amplitude of vibration displacement, μm
    0-125, 0-250 (DS-0 eddy-current transducer)
    0-100, 0-250 (DS-1 eddy-current transducer)
    0-500, 0-1 000 (DS-2 eddy-current transducer)
    0-2 000 (DS-3 eddy-current transducer)
  • Operating frequency range, Hz
    2-500
  • Installation gap (default), mm
    1.0±0.2 (DS-0 eddy-current transducer)
    1.0±0.2 (DS-1 eddy-current transducer)
    1.5±0.2 (DS-2 eddy-current transducer)
    5.0±0.2 (DS-3 eddy-current transducer)

Interface

  • Output signal type
    "current loop" 4-20 mA (for ver. A, B, E)
    0-10 V (for ver. E)
    digital output (for ver. E)
  • Supply voltage range, V
    10-24

Explosion protection

  • Kind
    intrinsically safe circuit
  • Marking
    0Ex ia IIC T6...T1 Ga X (DS-x eddy-current transducer)
    0Ex ia IIC T6...T1 Ga X (AS 1xx converter)
    0Ex ia IIC T6...T4 Ga X (connection box)
  • Connection
    through TIK-BIS safety barrier

Design parameters

  • Overall dimensions, mm
    D(6.8) L(50; 70; 90) (DS-0 eddy-current transducer)
    D(8.5) L(30; 50; 70; 90) (DS-1 eddy-current transducer)
    D(14.5) L(50; 70; 90) (DS-2 eddy-current transducer)
    D(22.5) L(90) (DS-3 eddy-current transducer)
    98.5x27.5x59 (AS 131 converter for ver. A)
    Ø30x66.5 (AS 1312 converter for ver. B)
    105x22.5x67 (AS 131 converter for ver. E)
    277x254.5x83 (connection box)
  • Weight, kg, not more than
    0.25 (DS-0, DS-1 ver. 00, 01 eddy-current transducer)
    0.35 (DS-0, DS-1 ver. 02 eddy-current transducer)
    0.3 (DS-2 ver. 00, 01 eddy-current transducer)
    0.4 (DS-2 ver. 02 eddy-current transducer)
    0.4 (DS-3 ver. 00, 01 eddy-current transducer)
    0.5 (DS-3 ver. 02 eddy-current transducer)
    0.1 (AS 131 / AS 1312 converter)
    2.5 (connection box)
  • Mounting type
    M8x1 threaded hole (DS-0 eddy-current transducer)
    M10x1 threaded hole (DS-1 eddy-current transducer)
    M16x1 threaded hole (DS-2 eddy-current transducer)
    M24x1 threaded hole (DS-3 eddy-current transducer)
  • Protection class
    IP65/IP68 (DS-x eddy-current transducer)
    IP20 (AS 1xx converter)
    IP65/IP67 (connection box)

Operating parameters

  • Operating temperature range, ºС
    -40...+135 (DS-x ver. 00, 01 eddy-current transducer)
    -40...+70 (DS-x ver. 02 eddy-current transducer)
    -60...+135 (DS-x ver. 00, 01 eddy-current transducer, climatic version HL)
    -60...+70 (DS-x ver. 02 eddy-current transducer, climatic version HL)
    -40...+70 (AS converter)
    -60...+70 (AS converter, climatic version HL)

Reliability parameters and manufacturer's guarantees

  • MTBF, hours, not less than
    40 000
  • Warranty service life, months
    24
  • Service life, years, not less than
    10
  • Installed device life, hours, not less than
    80 000
  • Time between verifications
    1 year
Service center